All Broadley-James dissolved oxygen sensors are available with optional right angle connector. This low profile design provides:
- Protection against cable fatigue
- Improved grip for sensor extraction
- Increased clearance between vessels
Specification
- 316L Stainless Steel
- Serialized for traceability
- RA32 finish with electropolish on wetted parts
- Internal and external silicone o-rings
- Secured retainer ring
The internal seals of an Oxyprobe are all mechanical seals using o-rings. There are no adhesives to leak or fail in process, making this a more rugged and reliable sensor. Replacing the cathode is also much easier and faster since there is no sealant to clean from the threads. No sealant also makes the job of replacing the cathode much faster and cleaner.
Membrane Cartridges
Single, Four and 16 piece Cartridge Kits