| Step 2
In one hand, grip the membrane cartridge, which has been properly filled with electrolyte solution, by the thumb and fingers while holding the sensor assembly with the other hand as shown. |
Caution:The use of protective glovesand eyewear is recommended throughout the membrane cartridge refill and installation procedures. |
| Step 3
Gently slide the membrane cartridge over the cathode / anode assembly until the cartridge threads seat against the mating threads of the sensor assembly. Thread the membrane cartridge onto the sensor assembly until there is a flush fitting between the cartridge sleeve and the tube. With a proper fit, the sealing o-ring should no longer be visible. |
Note: The use of protective gloves and eyewear is recommended throughout the membrane cartridge refill and installation procedures. |